24 Common MDR Deficiencies - Orthopaedic Implants & Surgical Instruments
Updated: Sep 25
Developed by Hands On MedTech Regulatory and Clinical Experts

Orthopaedic implants, single-use instruments, reusable instruments and surgical kits often face close notified body review. That is because safety and performance usually depend on several connected evidence areas, not one report in isolation.
A good MDR technical file needs to bring the full picture together. Device design, materials, mechanical performance, biological safety, cleaning, sterilisation, packaging, usability, clinical evaluation, PMS, PMCF, labelling and risk management all need to tell the same story.
Under the MDR, technical documentation should not read like a folder of separate reports. It should show, clearly and traceably:
What the device is
How it works
How it is manufactured and controlled
What claims are being made
How the applicable GSPRs have been met; and
How the manufacturer will continue to check safety and performance after the device is placed on the market.
This insight highlights our top reasons why orthopaedic technical files continue to attract questions under the MDR.
For orthopaedic manufacturers, deficiencies often appear when individual documents look acceptable on their own but do not work together as a submission package. For example, the clinical evaluation may not match the IFU. The GSPR checklist may list reports without explaining how they support conformity. The biological evaluation may not be backed by complete material data. Packaging validation may not explain the selected worst case. PMS and PMCF plans may be too broad for the device’s residual risks and clinical claims.
We’ve compiled the 24 most common MDR deficiencies to highlight where issues arise most frequently:
1. The device description is incomplete or unclear
The file does not fully define the device, device family, variants, sizes, accessories, kit configurations, packaging options, reusable components, single-use components or instruments used with the device.
For orthopaedic systems, notified bodies usually expect a clear explanation of how the implants, instruments, trays, delivery devices and accessories are used together during surgery.
2. The intended purpose changes across the file
The intended purpose, indications, target population, anatomical site, intended users, clinical benefits, contraindications, warnings and procedural limits are not consistent.
This can happen across the IFU, labels, CER, SSCP, risk file, PMS/PMCF plan, GSPR checklist, implant card, patient information, surgical technique guide or marketing claims.
3. The qualification and classification rationale is too brief
The file does not explain clearly enough why the product is a medical device, accessory, implantable device, surgical instrument, reusable instrument or part of a system.
The classification rationale may not fully address invasiveness, contact duration, implant status, sterility, reuse, measuring function, active features or the device’s role in the overall procedure.
4. Device family grouping is not justified
Different sizes, models, materials, instruments, sterile configurations, generations or kit components are grouped together without enough explanation.
Notified bodies often question whether one evidence package can support all variants. The file should explain why the selected representative or worst-case configuration is appropriate, using technical, biological, clinical and manufacturing criteria.
5. Combination use is not controlled clearly enough
The file does not clearly state which implants, instruments, accessories, trays, delivery systems, third-party components, imaging methods or ancillary devices are required, optional or compatible.
For orthopaedic systems, it should be clear how safety and performance are maintained when components are used together.
6. Material and component specifications are incomplete
Raw materials, finished device materials, coatings, surface treatments, additives, colourants, processing aids, packaging materials and patient or user contacting components are not fully specified.
Common gaps include missing drawings, tolerances, surface finish requirements, supplier specifications, certificates, material grades or acceptance criteria.
7. Design and manufacturing information is too high-level
Manufacturing flowcharts, site responsibilities, subcontractor activities, special processes, environmental controls, in-process checks, incoming inspection criteria and final release tests do not give enough detail.
The assessor needs to understand how the device is made, which sites are involved, which steps are critical and how conformity is maintained.
8. Supplier and subcontractor controls are not traceable
Critical outsourced activities are not clearly linked to supplier qualification, technical agreements, specifications, purchasing controls and change-control requirements.
This may include machining, moulding, coating, cleaning, packaging, sterilisation, laboratory testing, warehousing or design support.
9. GSPR traceability is weak
The GSPR checklist lists standards or reports but does not explain applicability, non-applicability, partial applicability, the method used to demonstrate conformity or the exact controlled-document location of the evidence.
Broad references to whole reports often lead to further questions.
10. The standards and state of the art strategy is not current
The submission relies on outdated standards, with drawn test methods, legacy assumptions or previous notified body acceptance without a current state of the art assessment.
Where older standards or alternative methods are used, the file should include a gap assessment and a clear technical rationale.
11. Risk management is disconnected from the wider file
The risk management file does not clearly link hazards, foreseeable misuse, surgical workflow, residual risks, risk controls, verification of risk-control effectiveness, IFU warnings, usability evidence, clinical evaluation, PMS/PMCF data and benefit-risk conclusions.
In weaker files, risks are listed but do not shape the evidence strategy.
12. Benefit-risk conclusions are too generic
The file concludes that the benefit-risk profile is acceptable, but does not provide enough quantitative or clinically meaningful support.
For orthopaedic devices, the file should address complication rates, device deficiencies, revision or reoperation risks, residual risks, mechanical failure modes, adverse events, state of the art alternatives and the claimed clinical or procedural benefits.
13. Mechanical and functional performance evidence does not support all claims
Bench, laboratory, simulated-use or functional testing does not cover all relevant claims, indications, configurations or use conditions.
For orthopaedic implants and instruments, this may include fixation strength, insertion and removal, deployment, torque, pull-out, fatigue, wear, corrosion, particulate generation, cutting or drilling performance, handling, dimensional compatibility and procedure-specific performance.
14. Cleaning and contamination control evidence is under-developed
For implants and instruments, common gaps include incomplete identification of contaminants, residues, particles, endotoxin, bioburden, cleaning agents, intermediate cleaning steps, final cleaning steps, method validation, sampling rationale, acceptance criteria, routine monitoring and revalidation triggers.
This matters especially where machining, moulding, polishing, coating or outsourced processing is involved.
15. Verification and validation reports are listed but not critically evaluated
Reports are included, but the file does not explain what was tested, why the method was suitable, whether deviations occurred, how acceptance criteria were selected, whether results support the GSPRs and whether the evidence applies to all marketed variants.
16. The biological evaluation is not device-specific enough
The biological evaluation does not fully consider contact type and duration, cumulative exposure, materials, processing residues, cleaning agents, packaging interaction, sterilisation effects, degradation, wear debris, extractables and leachables, target population, toxicological risk, evaluator competence or post-market biological safety signals.
17. Worst-case test samples are not justified
The tested device, lot, size, geometry, material, surface finish, manufacturing route, sterilisation status, ageing condition, reprocessing cycle or packaging configuration is not clearly representative of the marketed device.
Notified bodies often challenge testing where the selected sample is not justified as the right worst case.
18. Sterilisation validation is not clearly linked to the device family
The submission does not adequately justify adoption into an existing sterilisation family or load configuration.
Common gaps include incomplete protocols or reports, missing raw data, weak product compatibility rationale, unclear SAL justification, insufficient bioburden strategy, inadequate EO residuals or radiation dose rationale, and poor linkage between sterilisation, packaging and shelf-life evidence.
19. Packaging and sterile barrier validation does not reflect the marketed configuration
The sterile barrier system is not clearly defined, or the file does not explain single versus double sterile barrier claims.
Deficiencies often relate to packaging specifications, seal parameters, OQ/PQ rationale, equipment equivalence, sample size, transportation simulation, accelerated ageing, real-time ageing, labelling durability and post-ageing sterile barrier integrity.
20. Shelf-life and stability claims are not objectively demonstrated
The file claims a shelf-life without enough evidence that mechanical performance, packaging integrity, sterility, biocompatibility, labelling readability, material properties and usability remain acceptable for the full claimed period.
The storage and transport conditions should also be defined.
21. Labelling, IFU and surgical technique information do not align
Labels, IFU, surgical technique guides, implant cards, patient leaflets, SSCPs, training materials and promotional claims say different things.
Common gaps include missing target population, intended users, contraindications, residual risks, warnings, variant information, sterility status, single-use or reuse statements, MRI information where applicable, procedural limits and required accessories.
22. Usability and surgical workflow evidence is not sufficient
The file does not show that intended users can safely prepare, assemble, handle, implant, deliver, deploy, retrieve, clean, reprocess or dispose of the device.
Reliance on training, warnings or surgical technique instructions may be questioned if it is not supported by usability or human-factors evidence.
23. The clinical evaluation strategy does not fit the device type
The CEP and CER do not clearly justify the level of clinical evidence needed for the device.
This is especially important for implantable devices, higher-risk devices, novel designs, devices with limited market history, devices relying on indirect clinical benefit, or instruments that enable a specific implant procedure.
24. Equivalence and similar-device arguments are not well supported
Biological, technical and clinical equivalence may be asserted without enough detail, access to relevant technical data or comparison of differences.
Where differences exist, the file should explain why they do not affect clinical safety and performance. Similar-device benchmarking may also be incomplete or inconsistent with the state of the art and PMCF strategy.
Explore common MDR deficiencies in orthopaedic implants and surgical instruments - download your free copy of our report from the Resources page on our website.
If you have any questions, feel free to contact us for a 30-min Regulatory Clarity Call. We can discuss whether your file is likely to attract avoidable notified body questions; how well the IFU, CER, risk file, GSPR checklist, PMS, PMCF and SSCP align; whether your clinical evidence strategy fits the device type and claims; where biological evaluation, cleaning, sterilisation, packaging or shelf-life evidence may need stronger justification; and which practical steps to prioritise before submission or deficiency response.